• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 7MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 7MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72204395
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an unspecified surgery, the screw biosure broke.It is unknown if there was a back-up device available or if there was a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 7MM X 25MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18040802
Report Number1219602-2023-02120
Device Sequence Number20707828
Product Code MAI
UDI-Device Identifier00885554036718
UDI-Public00885554036718
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2025
Device Catalogue Number72204395
Device Lot Number51005124
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date10/31/2023
Supplement Report FDA Received Date01/11/2024
Date Device Manufactured02/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-