| Catalog Number |
386811 |
| Medical Device Problem Codes |
Leak/Splash (1354); Migration or Expulsion of Device (1395); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd cathena 20gx1.25in winged bc catheter backed out of vein the following information was provided by the initial reporter; pivc go out the vein after contrast injection (scanner) - permeability is ok before the exam after use consequences: extravasation of contrast medium.In some cases, radiology has been performed to visualize the extravasation.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on device history record review, no abnormality was observed during the production of the potential affected batches.Current control there is a daily outgoing and 2-hourly in-process visual inspection to detect catheter tubing and adapter damage, and daily outgoing visual inspection to detect cannula damage.There is also a daily outgoing functional test on cannula and catheter penetration force to detect cannula and catheter tip damage.As no sample is returned for evaluation, actual root cause could not be established.
|
| |
|
Event or Problem Description
|
|
Pivc go out the vein after contrast injection (scanner) - permeability is ok before the exam after use consequences: extravasation of contrast medium.In some cases, radiology has been performed to visualize the extravasation.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on device history record review, no abnormality was observed during the production of the potential affected batches.Current control there is a daily outgoing and 2-hourly in-process visual inspection to detect catheter tubing and adapter damage, and daily outgoing visual inspection to detect cannula damage.There is also a daily outgoing functional test on cannula and catheter penetration force to detect cannula and catheter tip damage.As no sample is returned for evaluation, actual root cause could not be established.H3 other text: see h10 manufacture narrative.
|
| |
|
Event or Problem Description
|
|
Pivc go out the vein after contrast injection (scanner) - permeability is ok before the exam.After use: consequences: extravasation of contrast medium.In some cases, radiology has been performed to visualize the extravasation.
|
| |
|