| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported a meterset unit (mu) discrepancy between beam monitor unit display module (bmdm) and mosaiq.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported a monitor unit (mu) discrepancy between beam monitor unit display module (bmdm) and mosaiq.It was ascertained from the software logs that an abnormal termination occurred on the linac.The linac reported the mu count from the beam 1 channel to mosaiq and the treatment resumed delivering the remaining mu.Mosaiq is working as designed and intended and based on the available information there has been no mistreatment.
|
| |
|