| Catalog Number |
0166V14S |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be user related (example: over aspirated, incorrect syringe/collapse lumen/sac close eye/valve damage).The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿structure & composition: the product include three series: bardex, bardic and bardia.The composition of each series is as follows: bardex, bardic and bardia bi-lumen series are composed of catheter body, inflation cone interface, balloon and valve.Bardia tri-lumen is composed to tube body, inflation cone interface, deflation cone interface, flush cone interface, balloon and valve.¿the product is with silicone coating.The material of catheter body, inflation conical interface, deflation conical interface, flushing conical interface, balloon is natural latex.The material of valve is polypropylene.Sterilized by radiation.Disposable.Scope of application: foley catheter is intended for use in the drainage of urine from the bladder of children and adults.Precaution: this product contains natural rubber latex which may cause allergic reaction.Sterile unless package has been opened or damage.Warning: do not use petroleum substrate lubricants with the latex- based urethral catheters such as petroleum jelly and label liquid paraffin, which will damage latex and may burst balloon.Water-soluble lubricant can be used.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gentle insert a syringe in the catheter valve.Never use excessive force to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital practical, the valve may be cut off.If this fails, please contact an adequately trained professional for assistance, as directed by hospital practical.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: catheters need to be stored at room temperature and keep away from the direct light and preferably stored in the original packing box.Do not use the product of have latex allergy¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the nurse gave the patient a preoperative catheterization on september 2nd, 2023, before the operation started.After washing their hands, the nurse tested the catheter leakage and found that the balloon could not be withdrawn.They replaced the catheter immediately.It was impossible to pull out the urinary catheter.As per event cause analysis and description, stated that the water injection interface of the urinary catheter slipped into the y-shaped opening of the urinary catheter, making it impossible to withdraw the fixed liquid.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is inconclusive due to condition of the sample received.Based on the attached photo, it was observed the catheter was still inflated.However, the reported event could not be confirmed due to poor sample condition.A potential root cause for this failure could be pinch lumen/collapse lumen/thick sac thickness/valve damage/short inflation lumen.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "structure and composition: the product include three series: bardex, bardic and bardia.The composition of each series is as follows: bardex, bardic and bardia bi-lumen series are composed of catheter body, inflation cone interface, balloon and valve.Bardia tri-lumen is composed to tube body, inflation cone interface, deflation cone interface, flush cone interface, balloon and valve.¿the product is with silicone coating.The material of catheter body, inflation conical interface, deflation conical interface, flushing conical interface, balloon is natural latex.The material of valve is polypropylene.Sterilized by gamma radiation.Disposable.Scope of application: foley catheter is intended for use in the drainage of urine from the bladder of children and adults.Precaution: this product contains natural rubber latex which may cause allergic reaction.Sterile unless package has been opened or damage.Warning: do not use petroleum substrate lubricants with the latex- based urethral catheters such as petroleum jelly and label liquid paraffin, which will damage latex and may burst balloon.Water-soluble lubricant can be used.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gentle insert a syringe in the catheter valve.Never use excessive force to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital practical, the valve may be cut off.If this fails, please contact an adequately trained professional for assistance, as directed by hospital practical.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: catheters need to be stored at room temperature and keep away from the direct light and preferably stored in the original packing box.Do not use the product of have latex allergy" correction: d, h this report references a us equivalent device.The us udi equivalent for this product number is used h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the nurse gave the patient a preoperative catheterization on september 2nd, 2023, before the operation started.After washing their hands, the nurse tested the catheter leakage and found that the balloon could not be withdrawn.They replaced the catheter immediately.It was impossible to pull out the urinary catheter.As per event cause analysis and description, stated that the water injection interface of the urinary catheter slipped into the y-shaped opening of the urinary catheter, making it impossible to withdraw the fixed liquid.
|
| |
|