| Model Number |
3280 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.The target lesion was located in the left circumflex artery (lcx).A rotalink advancer and a 1.25mm rotalink burr were selected for use.During the procedure, it was noted that the burr and the steel wire of the advancer were stuck together at high speed.However, as another advancer was not available, the procedure could not be continued.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The advancer, drive shaft, and handshake connection were visually and microscopically examined.Inspection of the device found that the handshake connection was bent.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.The target lesion was located in the left circumflex artery (lcx).A rotalink advancer and a 1.25mm rotalink burr were selected for use.During the procedure, it was noted that the burr and the steel wire of the advancer were stuck together at high speed.However, as another advancer was not available, the procedure could not be continued.There were no patient complications reported.
|
| |
|