| Model Number |
100/897/070 |
| Medical Device Problem Codes |
Disconnection (1171); Loose or Intermittent Connection (1371); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tracheostomy tube is not being held securely by the flange.Despite being locked, the tube slipped through and advanced 3.5cm into the patient's airway.Also noted that the connector at the end of the tracheostomy tube (to connect to circuit) is loose and disconnects easily.No injury reported.Patient involvement.Patient information is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device sample was received in used condition without the original packaging.Per visual inspection it was noticed that blue locking lever is incorrectly assembled with flange body (it cannot lock).Disassembled locking lever and correctly assembled it back.We can confirm that after assembly locking mechanism works well.The complaint was confirmed.The root cause was identified to be in design of adjustable flange which allows to unscrew blue locking lever.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.Similar customer complaints were received in the past therefore this issue was escalated to complaint trend process.The occurrence of this issue is being monitored.
|
| |
|
Search Alerts/Recalls
|