| Model Number |
101/860/060 |
| Medical Device Problem Codes |
Disconnection (1171); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: month and year of event have been provided, day is unknown.Lot number, expiration date and manufacture date are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that after 10 days in use for the patient, the pilot balloon had detached from the inflation tube connection.When attempting to measure cuff pressure, the pilot balloon detached from the inflation tube connection, even though no significant force was applied.Adverse effects are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: regulatory.Responses@icumed.Com.One device was received.Per visual inspection, it was noticed that inflation line was detached from pilot balloon.The complaint was confirmed.A root cause for the condition has not been determined.A device history record (dhr) review could not be performed as the lot number was unknown.This issue was escalated with further actions being taken accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: insufficient information was provided to be able to determine the full udi.**udi related data quality updates only**.
|
| |
|