• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR BAREWIRE EMBOSHIELD FILTER DELIVERY WIRE; EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR BAREWIRE EMBOSHIELD FILTER DELIVERY WIRE; EMBOLIC PROTECTION SYSTEM Back to Search Results
Catalog Number 22440-31
Medical Device Problem Codes Difficult to Remove (1528); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion in the superficial femoral artery (sfa) with 95% stenosis, mild calcification and mild tortuosity.The 315 barewire workhorse guidewire and the atherectomy device got stuck with each other, making it difficult to withdraw the atherectomy device.After various attempts to successfully withdraw the atherectomy device, it was removed independently.However, when preparing to retrieve the emboshield nav6 filter from the popliteal artery, the retrieval catheter could not be delivered.It was noted that the barewire portion that was outside the anatomy was bent.So the physician used the 5f single-bend catheter to retrieve the filter.There was no adverse patient sequelae and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned product.The reported difficult to advance and difficult to remove were unable to be confirmed due to the condition of the returned unit.The reported bends on the barewire were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other complaints reported from this lot.The investigation determined that the reported difficulties were likely due to circumstances of the procedure.It is likely that the clearance between the inner diameter of the atherectomy device and outer diameter of the barewire became reduced during use, resulting in resistance and the devices becoming stuck together causing difficulty to remove.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAREWIRE EMBOSHIELD FILTER DELIVERY WIRE
Common Device Name
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18048779
Report Number2024168-2023-12101
Device Sequence Number15472929
Product Code NTE
UDI-Device Identifier08717648137808
UDI-Public08717648137808
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue Number22440-31
Device Lot Number3041061
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/10/2023
Supplement Date Received by Manufacturer11/30/2023
Initial Report FDA Received Date11/01/2023
Supplement Report FDA Received Date12/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMBOSHIELD NAV6 FILTER.; ROTAREX.
Patient SexUnknown
-
-