Model Number N/A
Medical Device Problem Codes Difficult to Insert (1316); Dent in Material (2526)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: foreign - (b)(6).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2023-00299.
 
Event or Problem Description
It was reported that during a knee arthroplasty, the mating slot on the front of the tibial baseplate broke while engaging the articular surface via persona articular surface inserter.The articular surface failed to engage in an optimal position.It was stuck and surgeon struggled to remove it out, resulting in a dent and scratch on the articular surface.Therefore, a new articular surface was opened as per request.The broken tibial baseplate component was already cemented into the tibia so it remained implanted.The surgical delay was approximately 20 minutes to remove the stuck articular surface and to put in a new one.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned articular surface lot#: 65739545 exhibits signs of use (flared dovetail) and damage as alleged.Tibial implant was not returned and no photos provided to confirm fracture.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined for the mating issue or the broken tibial plate.However, leaving a broken/damaged tibial plate in the patient is considered use error as the ifu states 'do not use any component if damage is found or caused during setup or insertion.' this complaint has been confirmed for the mating issue via the flared articular surface but the tibial plate fracturing could not be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING (PS) RIGHT 10 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18048825
Report Number3007963827-2023-00300
Device Sequence Number20839684
Product Code MBH
UDI-Device Identifier00889024243750
UDI-Public(01)00889024243750(17)280331(10)65739545
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42522400710
Device Lot Number65739545
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2023
Supplement Date Received by Manufacturer04/19/2024
Initial Report FDA Received Date11/01/2023
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-