MAUDE Adverse Event Report: ACCUTRON, INC. ULTRA PC% FLOWMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCUTRON, INC. ULTRA PC% FLOWMETER Back to Search Results
Model Number 31975
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vomiting (2144)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Accutron requested and received the unit subject of the event for evaluation.The unit was tested and was found to be operating according to specifications.The ultra pc% flowmeter user manual states (2), "patient should always be closely monitored during nitrous oxide use.If patient has an adverse reaction, reduce or stop the flow of nitrous oxide as needed.The o2 flush button can be used to rapidly purge the lines of n2o.If patient does not show signs of quick recovery, remove nasal mask and treat with pure oxygen from either the o2 resuscitator fitting or an auxiliary oxygen tank using a demand valve, oxygen assisted manual resuscitator, or equivalent.Call for emergency assistance if rapid response is not achieved." no additional issues have been reported.
 
Event or Problem Description
The user facility reported that during patient procedures involving the ultra pc % flowmeter, three patients woke up and began to vomit.No additional medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA PC% FLOWMETER
Common Device Name
FLOWMETER
Manufacturer (Section D)
ACCUTRON, INC.
1733 west parkside ln.
phoenix AZ 85027
Manufacturer (Section G)
ACCUTRON, INC.
1733 west parkside ln.
phoenix AZ 85027
Manufacturer Contact
daniel davy
1733 west parkside ln.
phoenix, AZ 85027
4403927453
MDR Report Key18049724
Report Number2020813-2023-00004
Device Sequence Number7465656
Product Code BSZ
UDI-Device Identifier00813830021158
UDI-Public(01)00813830021158
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K970163
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number31975
Device Catalogue Number31975
Device Lot Number(10)2637971
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/01/2023
Supplement Report FDA Received Date06/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-