| Model Number |
D401 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, this implantable cardioverter defibrillator (icd) was found to be in safety mode.The field representative will discuss with the physician about the replacement, and it will be sent for analysis.This device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that, this implantable cardioverter defibrillator (icd) was found to be in safety mode.The field representative will discuss with the physician about the replacement, and it will be sent for analysis.This device remains in service.No adverse patient effects were reported.Additional information was received which indicated that, this implantable cardioverter defibrillator (icd) was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that both brady and tachy therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing but was likely the result of exposure to radiation, either therapeutic or environmental.
|
| |
|
Event or Problem Description
|
|
It was reported that, this implantable cardioverter defibrillator (icd) was found to be in safety mode.The field representative will discuss with the physician about the replacement, and it will be sent for analysis.This device remains in service.No adverse patient effects were reported.Additional information was received which indicated that, this implantable cardioverter defibrillator (icd) was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|