| Catalog Number |
68LP-325 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Dhr trend summary: considering that there is not an adverse trend for this type of event no additional actions are deemed required at this time.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: a review of the device noted one opening on the posterior side, assessed as a fold flaw opening.Additional observations noted crease and deformation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.The device has been explanted.
|
| |
|