NORTHGATE TECHNOLOGIES INC. CONMED GS2000; 50L ABDOMINAL INSUFFLATOR Back to Search Results
Model Number 72-00258-0
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Codes Abdominal Distention (2601); Insufficient Information (4580)
Date of Event 09/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On 02oct2023 northgate technologies was made aware of the following alleged event " the patient was injured however hospital contact would not provide specifics." "device monitor displayed -9mmhg however the patient was clearly hyper-distended.Procedure stopped and the device was reset [?][?][?][?]the same issue happened again.Theatre tech swapped out the unit for another gs2000 however this didn't operate correctly either [redacted hospital contact] believes this may have been due to incorrect setup of gas supply but will be returning [the device] for inspection regardless." the event occurred in australia.
 
Additional Manufacturer Narrative
On 02oct2023 northgate technologies, inc was made aware of the alleged event that occurred in australia.Based on the information provided through 01nov2023, there has been a report of a patient injury and possible malfunction of the device pressure reading, however despite multiple attempts to get further clarification around the type and nature of the injury and usage of the device no further information is available.The reported device is being returned for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.The device history record for (b)(6) from february of 2014 (mo 11317) was reviewed and the device passed all testing.Nothing out of the ordinary was noted.The device has been returned to nti for repair / evaluation previously for software upgrade under rma 31128 ((b)(6) 2018) and the unit required replacement of proportional valve and valve controller pcb.The unit was recalibrated after software upgrades and passed all fqc testing.The unit was updated from rev c to rev e.The device was returned to nti via rma 33951 ((b)(6) 2022) with complaint of device not giving correct reading.Complaint was not verified device performed as expected.However, the device memory indicated f12 (pressure fault) as occurring once.Fault code #12 was not reproduced.Minor cosmetic damage was noted.Repairs were completed by replacing the bezel, overlay, touchscreen.The device was recalibrated and passed all fqc inspections and testing.
 
Additional Manufacturer Narrative
On (b)(6) 2023 northgate technologies, inc was made aware of the alleged event that occurred in australia.Based on the information provided through (b)(6) 2023, there has been a report of a patient injury and possible malfunction of the device pressure reading, however despite multiple attempts to get further clarification around the type and nature of the injury and usage of the device no further information is available.This issue will continue to be monitored through the complaint system to assure patient safety.The device history record for (b)(6) from (b)(6) 2014 ((b)(6)) was reviewed and the device passed all testing.Nothing out of the ordinary was noted.The device has been returned to nti for repair / evaluation previously for software upgrade under rma 31128 ((b)(6) 2018) and the unit required replacement of proportional valve and valve controller pcb.The unit was recalibrated after software upgrades and passed all fqc testing.The unit was updated from rev c to rev e.The device was returned to nti via rma 33951 ((b)(6) 2022) with complaint of device not giving correct reading.Complaint was not verified device performed as expected.However, the device memory indicated f12 (pressure fault) as occurring once.Fault code #12 was not reproduced.Minor cosmetic damage was noted.Repairs were completed by replacing the bezel, overlay, touchscreen.The device was recalibrated and passed all fqc inspections and testing.The device was received at nti on (b)(6) 2023.The device evaluation did not find any malfunctions that would have caused or contributed to a hyperdistended abdomen.The pressure sensing and venting safety features in the device were tested and operated as designed.The items noted during service (e.G., fault code 10, low flow at 5 lpm, regulator debris, regulator pressure relief valve venting early) would not have caused or contributed to the complaint event.The insufflator has 2 main safety mechanisms the first is the electronic (software) controlled pressure relief mechanism which was designed to ensure pressure relief at max pressure or when patient pressure exceeds set pressure by more than 3 mmhg for more than 5 sec.This is what is referred to in the operators manual and is dependent upon the set pressure of the device and the magnitude and duration of the over pressure condition.The second safety mechanism is a mechanical pressure relief valve that is intended to ensure that pressure overshoot does not exceed 45 mmhg for more than 15 seconds when establishing pneumoperitoneum.These were operating as designed and intended when tested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONMED GS2000
Common Device Name
50L ABDOMINAL INSUFFLATOR
Manufacturer (Section D)
NORTHGATE TECHNOLOGIES INC.
1591 scottsdale court
elgin IL 60123
Manufacturer (Section G)
NORTHGATE TECHNOLOGIES INC.
1591 scottsdale court
elgin IL 60123
Manufacturer Contact
todd gatto
1591 scottsdale court
elgin, IL 60123
MDR Report Key18051936
Report Number0001450997-2023-00006
Device Sequence Number18149257
Product Code HIF
UDI-Device Identifier20845854045395
UDI-Public20845854045395
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K120151
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number72-00258-0
Device Catalogue NumberGS2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2023
Supplement Date Received by Manufacturer11/07/2023
Initial Report FDA Received Date11/01/2023
Supplement Report FDA Received Date12/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-