|
An event regarding altr/abnormal ion level involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to pain, elevated cobalt and chromium levels, and altr.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.Although the reported lot code is in the scope of the capa, the reported failure mode is not in the scope of this recall.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.H3 other text : not returned to the manufacturer.
|
|
It is reported by counsel that claimant underwent right hip surgery on (b)(6) 2010, and was implanted with a restoration modular hip system and a lfit v40 cocr femoral head.He was revised on (b)(6) 2022, due to pain, elevated cobalt and chromium levels and altr.
|