• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM*STM ST,30+4L NK,16X11X150; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM*STM ST,30+4L NK,16X11X150; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 563516
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Osteolysis (2377); Inadequate Osseointegration (2646)
Date of Event 10/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #
=
>(b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary
=
> no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision due to osteolysis and aseptic loosening.Date of implant: (b)(6) 2000 date of revision: (b)(6) 2023 (left hip) treatment: revision; head, liner and cup were revised depuy products used: catalog id: 550514 lot number/product identifier: sb109322 component type: sleeve description: srom sleeve proximal titanium 16d lg catalog id: 560029 lot number/product identifier: 686078 component type: head description: srom head zirconia 28mm +6 catalog id: 563516 lot number/product identifier: 604035 component type: stem description: srom stem std 30mm +4l neck 16mm x 11mm x 150mm competitor cup, liner, and screws were used - were depuy components removed: yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18054744
Report Number1818910-2023-22304
Device Sequence Number7017822
Product Code JDI
UDI-Device Identifier10603295178194
UDI-Public10603295178194
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2000
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2004
Device Catalogue Number563516
Device Lot Number604035
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2023
Initial Report FDA Received Date11/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/1999
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMPETITOR CUP (SMITH & NEPHEW); COMPETITOR LINER (SMITH & NEPHEW); COMPETITOR SCREWS (SMITH & NEPHEW); S-ROM*HEAD FEMRL ZIRCONIA 28+6; S-ROM*SLEEVE PRX ZTT, 16D-LRG
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient Weight97 KG
-
-