• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DISP KIT, CURVED, KNOTLESS HIP FIBERTAK; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DISP KIT, CURVED, KNOTLESS HIP FIBERTAK; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number DISP KIT, CURVED, KNOTLESS HIP FIBERTAK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 10/4/2023, it was reported by a sales representative via phone that an ar-3636h self bunching kl 1.8 fibertak, hip, and an ar-3638dhc disposables kit for knotless hip fibertak suture anchor, curved had an issue.During a hip labral repair procedure on (b)(6) 2023, when the surgeon was trying to put the 4th anchor in, she drilled as per the required technique and felt resistance when inserting the anchor.When using a mallet to insert the anchor, a part of the metal insertion shaft broke off inside the patient's bone.The broken piece was able to be removed by 90%, but the distal end remained in the bone.Another larger anchor from a competitor was used to complete the procedure successfully with no harm to the patient.There was a case delay of 30 minutes, and no additional anesthesia was administered.After the procedure, it was noticed that the curved guide looked slightly bent, and the drill wasn't entering through the device as it normally did.Nothing had broken off the curved guide.Both devices were available for return.No pictures were taken.
 
Additional Manufacturer Narrative
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force and/or prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP KIT, CURVED, KNOTLESS HIP FIBERTAK
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18055661
Report Number1220246-2023-08520
Device Sequence Number5913961
Product Code LXH
UDI-Device Identifier00888867295544
UDI-Public00888867295544
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDISP KIT, CURVED, KNOTLESS HIP FIBERTAK
Device Catalogue NumberAR-3638DHC
Device Lot Number15078448
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer12/20/2024
Initial Report FDA Received Date11/02/2023
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-