• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; ARTHROSCOPE Back to Search Results
Catalog Number 284580
Medical Device Problem Codes Improper Flow or Infusion (2954); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.E3: reporter is a j&j sales representative.E3: reporter is a j&j employee.H4: the device manufacture date is currently unavailable.Udi: (b)(4).Investigation summary : the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: faulty main board; therefore, the unit is unrepairable.Per service reports, this complaint can be confirmed.The complete device was replaced to resolve the issues as system was deemed not repairable.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review : manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Event or Problem Description
It was reported by the service center in south africa that during preventative maintenance, it was discovered that the 4580 fms duo+ pump/shaver combo device had an inadequate pressure.The event was not related to surgery.There was no procedure nor patient involvement reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18056616
Report Number1221934-2023-03992
Device Sequence Number20541485
Product Code HRX
UDI-Device Identifier10886705020607
UDI-Public10886705020607
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K954465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2023
Initial Report FDA Received Date11/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-