MAUDE Adverse Event Report: MPRI CAPSUREFIX NOVUS LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSUREFIX NOVUS LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5076-35
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Code Asystole (4442)
Date of Event 10/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient was noted to have experienced asystole.   during use of the cardiac resynchronization therapy defibrillator (crt-d) a pacing problem, loss of capture occurred with probable connector, header issue.  the crt-d was explanted and replaced  the right ventricular (rv) lead exhibited pacer spikes, loss of capture, intermittent capture and a dislodge confirmed.  the rv lead was re-positioned and remains in use.  the right atrial (ra) lead noted with pacer spikes, loss of capture.  the ra lead remains in use.  no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 1258t lead implanted: (b)(6) 2014; dtbb1d4 crt-d implanted: (b)(6) 2018.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was reported that the patient experienced asystole.A pacing problem and loss of capture occurred with probable connector, header issue on the cardiac resynchronization therapy defibrillator (crt-d). the right ventricular (rv) lead showed periods of pacer spikes with no capture and intermittent capture.A lead dislodgement was confirmed. manual rv lead threshold testing was variable.The right atrial (ra) lead showed periods of pacer spikes with no capture.The crt-d was explanted and replaced.The rv lead was re-positioned and remains in use. the ra lead remains in use.  no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
**udi related data quality updates only** corrected: d4 (provide complete udi).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS LEAD MRI SURESCAN
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18059946
Report Number2649622-2023-30530
Device Sequence Number20511576
Product Code DTB
UDI-Device Identifier00681490729963
UDI-Public(01)00681490729963(17)120225(2
Combination Product (Y/N)Y
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/25/2012
Device Model Number5076-35
Device Catalogue Number5076-35
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 10/11/2023
Supplement Date Received by Manufacturer11/06/2023
08/01/2024
Initial Report FDA Received Date11/02/2023
Supplement Report FDA Received Date11/06/2023
09/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M72 LEAD, 5867-3M ADAPTOR.
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-