| Catalog Number |
SHPB545 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a female patient underwent an unspecified breast augmentation with a mentor memorygel, boost 545cc silicone breast implant experienced bilateral capsular contracture grade iii postoperative.Patient reports extreme pain daily due to capsular contractions.Patient reports her breast feel tight and uncomfortable.As a result, patient scheduled for bilateral removal on (b)(6) 2023.This report relates to the left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture grade iii.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received on november 9, 2023 indicated that the capsular grade for both sides were grade iv, and patient also underwent bilateral partial capsulectomy.The inspection of both breasts indicated: right breast inspection: baker grade iv capsule (palpable and visible deformity with associated pain) left breast inspection: baker grade iv capsule (palpable and visible deformity with associated pain) breast palpation: right breast palpation: implant malposition (superior) left breast palpation: implant malposition (superior).The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per additional information received on november 7, 2023 indicates that the patient underwent bilateral replacements as follow: l - sn: (b)(6), r - sn: (b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on december 01 , 2023: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the smo ro high pro boost gel545cc returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.S manufacturer¿s reference number: (b)(4).
|
| |
|