| Catalog Number |
320-15-05 |
| Medical Device Problem Codes |
Structural Problem (2506); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
10/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this 72 y/o male patient's right shoulder was revised to a reverse shoulder.The glenosphere locking screw and glenosphere appeared to have backed away from the glenoid baseplate.The glenosphere locking screw was sitting proud and ended up wearing through the liner and made contact with the humeral tray.Surgeon replaced the 38 glenosphere, glenosphere locking screw, liner, reverse shoulder torque screw and humeral tray with new components.Patient was last known to be in stable condition following the event.Photos attached.Unable to obtain x-rays.Devices are not returning.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) pending evaluation (d10) concomitant device(s): glenosphere locking screw, liner, reverse shoulder torque screw, humeral tray.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d4, d6b - explant date is unknown.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected/ additional information.The following sections were corrected/added: b1, b2, b5, d1, d8, g1, h6: clinical code, medical device problem code, component code, type of investigation, investigation findings and investigation conclusions.The revision reported was likely the result of the glenosphere locking screw backing out from the glenoid baseplate, which led to articulation between the screw and the humeral liner and humeral tray.This unintended contact likely led to wear of the liner and the tray.However, the sequence of events cannot be confirmed, and the cause of the screw loosening cannot be determined as the devices were not returned for evaluation and radiographs were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that this male patient's right shoulder was revised to a reverse shoulder.The glenosphere locking screw and glenosphere appeared to have backed away from the glenoid baseplate.The glenosphere locking screw was sitting proud and ended up wearing through the liner and made contact with the humeral tray.Surgeon replaced the 38 glenosphere, glenosphere locking screw, liner, reverse shoulder torque screw and humeral tray with new components.Patient was last known to be in stable condition following the event.Photos attached.Unable to obtain x-rays.Devices are not returning.
|
| |
|