• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD NANO¿ 2ND GEN PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD NANO¿ 2ND GEN PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 320122
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd nano¿ 2nd gen pen needle was unable to prime.Report 1 of 2.The following was received from the initial reporter: consumer reported during the flow check of last 2 boxes 320122 found 10-15 pen needles each box that clogged.Informed caller of proper non patient end placement.
 
Additional Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd nano¿ 2nd gen pen needle was unable to prime.Report 1 of 2.The following was received from the initial reporter: consumer reported during the flow check of last 2 boxes 320122 found 10-15 pen needles each box that clogged.Informed caller of proper non patient end placement.
 
Additional Manufacturer Narrative
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NANO¿ 2ND GEN PEN NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key18061272
Report Number2243072-2023-01988
Device Sequence Number6590983
Product Code FMI
UDI-Device Identifier10885403928659
UDI-Public(01)10885403928659
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K213478
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number320122
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2023
Supplement Date Received by Manufacturer11/30/2023
Initial Report FDA Received Date11/02/2023
Supplement Report FDA Received Date12/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-