Catalog Number A1108
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported that it was not possible to fix the mayfield triad skull clamp (a1108) to the patient's skull, which caused a skin laceration to the scalp/skull.A replacement device was used to complete the case.There was increased surgery time of 15/20 minutes.The patient is an adult and is reported as "ok.".
 
Additional Manufacturer Narrative
The mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the reported complaint was confirmed from the evaluation.The ratchet casting was bent due to misuse and could not be inserted smoothly into the skull clamp base.The received torque screw was not an integra product.There was movement in the lock and the unit required replacement of worn components.To resolve all issues, the internal parts to adjust the unit were replaced according to the manufacturer and the skull clamp's swivel lock assembly was cleaned.After completing the reassembly, the skull clamp was subjected to a successful function test.Root cause: complaint is confirmed via inspection of the unit.Probable root cause of the reported complaint is rough handling of the skull clamp and improper or suboptimal placement of the skull clamp on the patient.No corrective action has been initiated for the same or similar issue.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD TRIAD SKULL CLAMP
Common Device Name
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18062401
Report Number3004608878-2023-00199
Device Sequence Number7465768
Product Code HBL
UDI-Device Identifier10381780253518
UDI-Public10381780253518
Combination Product (Y/N)N
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberA1108
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/15/2023
Initial Report FDA Received Date11/02/2023
Supplement Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-