| Catalog Number |
8065751571 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported probe could not cut during vitrectomy surgery.A new product was changed to complete the surgery.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The opened probe returned for evaluation was not within the final lot number reported based on radio frequency identification (rfid) tag identification; therefore, no sample evaluation was performed.A review of the device history records traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Since the sample associated with the reported product and lot was not received at the manufacturing site, the root cause for the customer complaint issue cannot be determined.The returned sample was not within the final lot number based on rfid tag identification; therefore, specific actions with regard to this complaint cannot be taken.However, a non-conformance investigation has been completed in relation to the related non-conformances identified within the non-conformance review.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|