| Model Number |
9390A-1001 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via electrode pads.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device activity log.However, the device was put through extensive testing including functionality testing without duplicating the report.Internal inspection of the device found no discrepancies.No accessories was returned as part of this investigation.The main board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report updates information based upon your request which differs from the initial medwatch report filed.Please reference sections d1 updated, d4 catalog # removed, d4 primary udi # updated.
|
| |
|
Search Alerts/Recalls
|