• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX596T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx596t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 3 years and 3 months.Height: 93 centimeters.Weight: 14 kilograms.Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has indicated that the shunt assistant 2.0 has a blockage and the progav 2.0 valve and the ped.Pre-chamber are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The progav 2.0 valve is tested in the horizontal position.Because the shunt assistant 2.0 is not permeable, a computer controlled test is not possible.Accelerated outflow was detected in the progav 2.0.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in progav 2.0 and shunt assistant 2.0.To make the proteins / deposits in the shunt system more visible, they were colored using a staining solution.Results: based on our investigation results, we can determine an accelerated outflow and adjustment difficulties in the progav 2.0.The shunt assistant 2.0 has a blockage.The determined deposits can be named as the cause for these problems.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18066122
Report Number3004721439-2023-00335
Device Sequence Number7593621
Product Code JXG
UDI-Device Identifier04041906503536
UDI-Public4041906503536
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model NumberFX596T
Device Catalogue NumberFX596T
Device Lot Number20050513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2023
Initial Report FDA Received Date11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 YR
Patient SexFemale
Patient Weight14 KG
-
-