BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; TRANSPORT CULTURE MEDIUM Back to Search Results
Catalog Number 364956
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 10/04/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.D.4.Medical device expiration date: unknown.H.4.Device manufacture date: unknown.
 
Event or Problem Description
It was reported when using an bd vacutainer® urine complete cup kit during the process of preparing or completing transfer to the collection tube nurses are accidently inserting their finger in the recessed sharps area, which caused a needlestick injury.The following information was provided by the initial reporter.The customer stated: -customer problem: customer reports needlestick injuries while using bd urine collection cups.-steps taken with customer/troubleshooting: per pir3068079, customer states nursing staff is experiencing needlesticks with the urine collection cups.Per customer, in the process of preparing or completing transfer to the collection tube nurses are accidently inserting their finger in the recessed sharps area.Two different product numbers were provided, 364956 and 364975.
 
Event or Problem Description
It was reported when using an bd vacutainer® urine complete cup kit during the process of preparing or completing transfer to the collection tube nurses are accidently inserting their finger in the recessed sharps area, which caused a needlestick injury.The following information was provided by the initial reporter.The customer stated: -customer problem: customer reports needlestick injuries while using bd urine collection cups.-steps taken with customer/troubleshooting: per pir3068079, customer states nursing staff is experiencing needlesticks with the urine collection cups.Per customer, in the process of preparing or completing transfer to the collection tube nurses are accidently inserting their finger in the recessed sharps area.Two different product numbers were provided, 364956 and 364975.
 
Additional Manufacturer Narrative
H.6 investigation summary: bd had not received samples or photos for evaluation.Additionally, bd was unable to determine the specific lot number associated with this complaint; therefore, a review of the device history record could not be conducted.This complaint is unable to be confirmed.If additional information is made available, this complaint will be reopened to assess the level of investigation needed.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® URINE COMPLETE CUP KIT
Common Device Name
TRANSPORT CULTURE MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
jo doyka
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key18066342
Report Number1917413-2023-01114
Device Sequence Number7572613
Product Code JSM
UDI-Device Identifier50382903649568
UDI-Public(01)50382903649568
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number364956
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer11/07/2023
Initial Report FDA Received Date11/03/2023
Supplement Report FDA Received Date11/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-