• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH ENFIT SALEM SUMP PVC TUBE 18FR 48IN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENFIT SALEM SUMP PVC TUBE 18FR 48IN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8888264986E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Internal Organ Perforation (1987)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the tube was too rigid and did cause a gastric perforation; approximately 1-week post-installation, bile detected in abdominal drains.Upon surgical re-exploration, a perforation of the stomach was noted.Several attempts to gather information from the customer were made.To date, no response has been received.If additional pertinent information becomes available, the report will be updated.
 
Additional Manufacturer Narrative
The device history record review could not be performed since no lot number was provided.Because a sample was not returned, we were unable to perform functional and visual evaluations to confirm the reported condition and determine the root cause.If a sample is received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENFIT SALEM SUMP PVC TUBE 18FR 48IN
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key18066592
Report Number9612030-2023-03852
Device Sequence Number18082396
Product Code PIF
UDI-Device Identifier10884521779457
UDI-Public10884521779457
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number8888264986E
Device Catalogue Number8888264986E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer10/24/2023
Initial Report FDA Received Date11/03/2023
Supplement Report FDA Received Date01/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-