ARTHREX, INC. DISSECTOR, CURVED 4.0MM X 13CM; SAW BLADE Back to Search Results
Model Number DISSECTOR, CURVED 4.0MM X 13CM
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 10/13/2023, it was reported by a sales representative via sems-06056928 that an ar-8400cds dissector was pushed down and the teeth were bent in, it would not rotate when trying to engage.This occurred during a case with no patient effect reported.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
Additional information: it was reported by a sales representative that this event occurred during a meniscus repair on (b)(6) 2023 where a 4.0 torpedo was used to complete the case.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional information from the field, arthrex concluded that the most likely cause.The most likely cause of the reported failure is misuse of the product caused by the surgeon using excessive force and bending the device.If the device is bent during use, the device may fail to operate as intended.The complaint allegation was not confirmed.The device was not received for evaluation, and no evidence of the failure was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISSECTOR, CURVED 4.0MM X 13CM
Common Device Name
SAW BLADE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18066788
Report Number1220246-2023-08541
Device Sequence Number18096557
Product Code GFA
UDI-Device Identifier00888867043497
UDI-Public00888867043497
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDISSECTOR, CURVED 4.0MM X 13CM
Device Catalogue NumberAR-8400CDS
Device Lot Number15121203
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2023
Supplement Date Received by Manufacturer02/18/2025
Initial Report FDA Received Date11/03/2023
Supplement Report FDA Received Date02/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-