• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR; THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR; THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 08496609190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys anti-tshr on a cobas 6000 e601 module.No questionable results were reported outside of the laboratory.The sample initially resulted in an anti-tshr value of > 40.00 iu/l with a data flag.The result did not match the patient's clinical diagnosis and the sample was repeated.The repeat result was 1.45 iu/l.
 
Additional Manufacturer Narrative
The serial number of the e601 analyzer is (b)(6).The pinch tubing was replaced.A probe was checked and it was ok.Precision studies were performed.The investigation is ongoing.
 
Additional Manufacturer Narrative
Quality controls were within range.The investigation determined the replacement of the pinch tubing resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-TSHR
Common Device Name
THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18067006
Report Number1823260-2023-03514
Device Sequence Number7665543
Product Code JZO
UDI-Device Identifier07613336160305
UDI-Public07613336160305
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K193313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08496609190
Device Lot Number654109
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/15/2023
Supplement Date Received by Manufacturer11/21/2023
Initial Report FDA Received Date11/03/2023
Supplement Report FDA Received Date11/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-