MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804350-38
Medical Device Problem Codes Difficult to Remove (1528); Device Damaged by Another Device (2915); Material Deformation (2976)
Health Effect - Clinical Codes Myocardial Infarction (1969); Thrombosis/Thrombus (4440)
Date of Event 10/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.Based on the reported information, it is likely that the guide wire interacted with the implanted 3.50x38 xience skypoint during removal causing the reported difficulty to remove and subsequent material deformation.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.The reported patient effects of thrombosis and myocardial infarction are listed in the xience skypoint everolimus eluting coronary stent systems instructions for use as a known patient effects of coronary stenting procedures there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified lad.Two 3.5x38mm and 2.75x38mm xience skypoint stents were implanted.During removal of unspecified guide wire it was noted to have met resistance with the proximal 3.5x38mm xience skypoint stent; however, no other complications were noted or issues.About 30 minutes later the patient suffered an myocardial infraction (mi) and it was confirmed via angiogram that thrombus was present only on the proximal xience skypoint stent not on the distal xience skypoint stent that was implanted.Angioplasty was performed to treat the thrombus and medication was given to treat the mi.It was noted that proximal stent might have become flared when it met resistance with the unspecified guidewire during the original procedure.There was no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18067102
Report Number2024168-2023-12239
Device Sequence Number15393399
Product Code NIQ
UDI-Device Identifier08717648233449
UDI-Public08717648233449
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/06/2026
Device Catalogue Number1804350-38
Device Lot Number3031441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2023
Initial Report FDA Received Date11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.75X38MM XIENCE SKYPOINT
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-