H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code b20 -the device remains implanted and was therefore not available for engineering evaluation by gore.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), potential adverse events that may occur and/or require intervention include, but are not limited to endoleaks, aneurysm enlargement, and reoperation.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following was reported to gore: on (b)(6) 2018, the patient had an endovascular aneurysm repair (evar) procedure utilizing a gore® excluder® aaa endoprosthesis (rlt261418 & pxc141400).During follow up imaging on (b)(6) 2022, revealed a type 1a endoleak with the sac measured 7 cm on non-con ct.During follow up cta on (b)(6) 2023, the sac measured 7.9 cm.Then another cta done on (b)(6) 2023 the sac measured 8.2 cm x 7.5 cm.On (b)(6) 2023, the patient was brought in for a reintervention endovascular procedure in zone 9 infrarenal, in which the physician extended proximally by adding the gore® excluder® conformable aaa endoprosthesis (aortic extender conformable -cxa260005) device followed by aptus endo anchoring.This did not resolve the endoleak as shown in angiogram.The procedure was well-tolerated and without complication.The physician will continue to monitor things for now.
|