Catalog Number QNR916
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information the tubing leaked.The device was explanted and replaced.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report, and capa.No ncs nor capas were identified.There were 3 complaints identified against this lot; however, with various failure modes.Therefore, this is not considered a trend.Devices met specification prior to release.
 
Event or Problem Description
According to the available information the tubing leaked.The device was explanted and replaced.
 
Additional Manufacturer Narrative
Titan otr pump and cylinders 1 and 2 were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.A separation, surrounded by abrasion, was noted on the longer exhaust tube of the pump.This is a site of leakage.Both cylinders were received with tow sutures still attached.No functional abnormalities were noted with cylinder 1 or cylinder 2.Based on examination of the returned product, it was concluded that the abrasion marks noted on both exhaust tubes and inlet tube of the pump matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo enough to cause a separation.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA   3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key18068337
Report Number2125050-2023-01352
Device Sequence Number12702464
Product Code FHW
UDI-Device Identifier05708932319744
UDI-Public05708932319744
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/21/2020
Device Catalogue NumberQNR916
Device Lot Number4684263_QNR9161022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/03/2023
Supplement Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TITAN INFLATABLE PENILE PROSTHESIS ACCESSORIES; TITAN INFLATABLE PENILE PROSTHESIS ACCESSORIES; TITAN INFLATABLE PENILE PROSTHESIS FAMILY; TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-