| Model Number |
RA002-5555SL |
| Medical Device Problem Codes |
Material Erosion (1214); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Metal Related Pathology (4530)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.A review of the device history record (dhr) confirmed the device met all required quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that upon removal it was noted that the patient had developed metallosis which was debrided during the procedure.There were no complication due to the observed metallosis.No additional information has been provided.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the rod was returned to nuvasive for evaluation.Upon visual inspection of the rod, scoring marks on the distraction rod were observed which are generally associated with incremental distractions.The presence of black debris was observed between the surfaces of the distraction rod and the housing tube.In-house x-rays revealed that the distraction rod and lead screw were not concentric with respect to the housing body.In addition, the distraction rod appeared to be contacting the inner surface of the housing tube, which is a possible reason for the score marks on that side of the distraction rod.X-rays also discovered a broken radial bearing.The rod was then sectioned, and the o-ring was examined.The o-ring was found torn and the housing tube¿s grove was worn down from the distraction rod rubbing on the inner surface.It is possible that the presence of external bending forces on the rod could have led to the scoring and by extension the generation of metal debris.These external bending forces could have also caused the distraction rod to have unequal compression on the internal o-ring, which can result in a torn seal.A compromised seal can lead to a broken bearings as bodily fluids ingress into the internal mechanism, increasing the friction on the bearing.Due to the broken bearing, functional testing was not possible.No patient x-rays or blood work results were provided, therefore, the reported metallosis could not be confirmed.However, wear debris was confirmed visually as it was found on the device upon return.The black debris that was found is likely titanium debris caused by friction during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h6 (health effect - impact code and device code).
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|