• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 3.25 X 8.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 3.25 X 8.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIFNT3285
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 06/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implant located in dental position number #16 failed due to peri-implantitis.The doctor reports in the per that the procedure was not concluded by placing another implant.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE TAPERED CERTAIN IMPLANT 3.25 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18071618
Report Number0001038806-2023-02112
Device Sequence Number7769349
Product Code DZE
UDI-Device Identifier00844868021896
UDI-Public(01)00844868021896(17)270719(10)2022070324
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 11/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberXIFNT3285
Device Lot Number2022070324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-