| Model Number |
IAV09008008P |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Thrombosis/Thrombus (4440); Insufficient Information (4580)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the index procedure one in.Pact av access balloon catheter was used to treat the anastomosis of the left arm.Approximately 16 months post the index procedure the patient underwent a revascularization procedure during which an in.Pact av access balloon catheter was used to treat the axillary vein and arterial inflow and proximal non cannulation zone of the left arm.Approximately 26 months post the index procedure and 9 months post the revascularization procedure the patient suffered from thrombosis of the arteriovenous graft at the trailing edge of the stent.The patient reported no thrill or bruit in the arteriovenous graft.A shuntagram showed thrombosis at the anastomosis at the trailing edge of a non mdt stent implanted prior to study enrollment.The enrollment lesion included a single lesion of the venous anastomosis and the trailing edge of the stent.Treatment was performed during the shuntagram.An angioplasty and thrombectomy was performed which was successful.The patient recovered.The sponsor commented that this was a target lesion revascularization.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Update received (b)(6) 2025: during the index procedure the left arm venous anastomosis was treated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|