• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNIVERS REVERS TRIAL 36 +3; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERS REVERS TRIAL 36 +3; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number UNIVERS REVERS TRIAL 36 +3
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during an univers revers shoulder prosthesis surgery the device produced debris.There was no harm for patient, operator or third party.The surgery was finished successfully and the same device was used anyway.It was not necessary to switch the surgical technique or do a second surgery.Update nroe 17-oct-2023: further information revealed that all broken-off pieces could be retrieved from the patient.
 
Additional Manufacturer Narrative
Additional information: h6.Complaint is confirmed.Visual evaluation notices that the edges around the device were damaged, and one of the borders of the hole was broken.No broken pieces were returned for investigation.Functional testing was not performed due to the damage to the device.Functional testing was not performed due to the damage to the device.The most likely cause of this failure is attributed to wear and tear damage incurred over repeated usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS REVERS TRIAL 36 +3
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18072460
Report Number1220246-2023-08560
Device Sequence Number12678883
Product Code LXH
UDI-Device Identifier00888867062146
UDI-Public00888867062146
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUNIVERS REVERS TRIAL 36 +3
Device Catalogue NumberAR-9530S-03
Device Lot Number250120600
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2023
Supplement Date Received by Manufacturer10/16/2023
Initial Report FDA Received Date11/06/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-