| Model Number |
N/A |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, that the unit did not mesh evenly and tore tissue.There was no reported harm, injury, or delay as there was no patient involvement.Due diligence is complete.No additional information is available.No adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: review of the most recent repair record determined the test cut passed and the reported issue could not be duplicated; however, the device was slightly out of calibration.The unit was disassembled, cleaned, and calibrated.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information is available regarding the incident.
|
| |
|