| Model Number |
799-EGL2-00-01 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Calibration Problem (2890); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed "compressor failure -1001," "compressor fault-2001," "self check failure -1003," and "runtime calibration failure - 1051" error messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.However, the device was put through extensive testing without duplicating the report.Internal inspection of the device found no discrepancies.The flow screens, smart pneumatic (spm) board, and spm tubing were replaced as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report updates information based upon your request which differs from the initial medwatch report filed.Please reference updated section d1 brand name, and section d4 catalog number that does not apply and should be removed.
|
| |
|