• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER VASCULAR PLUG 4; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER VASCULAR PLUG 4; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Catalog Number 9-AVP038-008
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event or Problem Description
It was reported that on (b)(6) 2023, an 8mm amplatzer vascular plug iv was chosen for procedure.During device preparation, it was noted there was a discrepancy between the device's outer packaging and inner packaging.On one label, it was stated the compatible catheter's inner diameter to be greater than or equal to 0.038in.On the other label, it stated the compatible catheter's inner diameter to be less than or equal to 0.038in.It was reported the device was still used in the procedure without additional issues or delay with a 0.038 inch inner diameter non-abbott diagnostic catheter.There was no report of any adverse patient consequences to the patient.
 
Additional Manufacturer Narrative
An event of mislabeling was confirmed.The investigation found the complaint is in scope of capa 128015 and the devices were determined to be used as is.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG 4
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18073417
Report Number2135147-2023-04870
Device Sequence Number7684861
Product Code KRD
UDI-Device Identifier00811806012384
UDI-Public00811806012384
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K113658
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9-AVP038-008
Device Lot Number8216968
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/12/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date11/06/2023
Supplement Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-