| Catalog Number |
620051180 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that after the case, the sterile processing tech noticed the humeral trial was still put together when it came to decon.When the tech tried to remove the offset taper she noticed that 2 of the tabs on the 51x18 trial broke off and went down the drain, it did break into multiple pieces.This happened post op.So there was no negative impact to the patient, nor was surgery time extended due to this.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added : d9 corrected: d4 primary udi number.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary it was reported that (b)(6) 2023 after the case.The sterile processing tech informed me that 2 of the humeral trials were still put together when they came to decon.When the tech tried to remove the offset taper she noticed that 2 of the tabs on the 51x18 (item# 6200-51-180, lot# 318533) broke off and went down the drain, it did break into multiple pieces but the tabs are not available since they went down the drain.She also said when she removed the adapter from the 51x20.5 (item# 6200-51-205, lot# 318534) she noticed 3 of the locking tabs were cracked.That trial is still in 1 piece.This happened post op.So there was no negative impact to the patient, nor was surgery time extended due to this.The product was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the humeral head trial 51x18 had two tabs broken.Fragments were not returned for evaluation.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces while disassembling the offset taper adapter trial, per inhance¿ shoulder system anatomic surgical technique with hybrid glenoid addendum (103832905 rev b) page 18, section humeral head trial disassembly: to remove the humeral head trial, assemble the head distractor to the impactor handle and place between the resection and the humeral head.To remove the offset taper adapter trial from the humeral head trial, rotate the offset taper adapter trial counterclockwise until the ramps unlock it from the head.Once the offset taper adapter trial reaches the e = eject position, it will engage the ramp and come out of the humeral head trial.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the humeral head trial 51x18 would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative:.
|
| |
|