|
|
| Model Number |
980U1ENDICU |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, while in use on a patient, this 980 ventilator generated a high priority alarm indicating that the ventilator was not ventilating and the main touch screen was not operable. the patient was removed from the ventilator and placed on an alternate ventilator with no injury reported.Although it was reported that the ventilator was not ventilating the patient, per code provided by the facility, the device entered into backup ventilation (buv) at the time of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the medtronic service engineer.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d8 updated section h6: evaluation code method add additional codes result code remove c20 and add c13 component code remove g07003 and add g07001 h3: medtronic conducted an investigation based upon all information received.It was reported that, while in use on a patient, this 980 ventilator generated a high priority alarm indicating that the ventilator was not ventilating and the main touch screen was not ope rable. although it was reported that the ventilator was not ventilating the patient, per code provided by the facility, the device entered into backup ventilation (buv) at the time of the reported event.The unit was not returned or made available to medtronic.The evaluation for the reported issue was performed by non-medtronic pers onnel.With the image provided, it was noted that the unit generated an ¿exhalation pressure reading out of range¿ code indicating that the unit entered into backup ventilation (buv).The customer performed the extended self test (est) and could not duplicate the reported issue.With the available information, the likely cause of the event was not determined the manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality s pecifications.The event is included in a trending and monitoring plan.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|