• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. MESHGRAFT II COMPLETE; SKIN GRAFT MESHER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. MESHGRAFT II COMPLETE; SKIN GRAFT MESHER Back to Search Results
Model Number N/A
Medical Device Problem Code Dent in Material (2526)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the test cut failed and the cutter and roller were damaged.The roller and cutter were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported the several grooves on feed roller were indented.There was no information provided regarding the timing of the event or if there was any patient involvement/harm that may have occurred.During device investigation, it was discovered that the unit failed the test cut.Due diligence is complete, no further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESHGRAFT II COMPLETE
Common Device Name
SKIN GRAFT MESHER
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18074957
Report Number0001526350-2023-01444
Device Sequence Number7769479
Product Code FZW
UDI-Device Identifier00889024372528
UDI-Public(01)00889024372528(11)180505(10)64025312
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00219500000
Device Lot Number64025312
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Initial Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-