• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. CHONDRAL PICK, CONCAVE, 90 DEG; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. CHONDRAL PICK, CONCAVE, 90 DEG; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number CHONDRAL PICK, CONCAVE, 90 DEG
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 10/10/2023, it was reported by a sales representative via (b)(4) that an ar-8656-02 chondral pick and an ar-8656-03 chondral pick were producing small metal shavings.This occurred during a case where the debris was retrieved.The devices were switched out and the case was completed.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 complaint is confirmed.Upon visual evaluation, the distal end of the internal curve of the shaft had an abrasion mark.The most likely cause can be attributed to user error of the device due to interference with another device, causing the damage to the shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHONDRAL PICK, CONCAVE, 90 DEG
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18075515
Report Number1220246-2023-08581
Device Sequence Number7665855
Product Code MDM
UDI-Device Identifier00888867307469
UDI-Public00888867307469
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCHONDRAL PICK, CONCAVE, 90 DEG
Device Catalogue NumberAR-8656-03
Device Lot Number1392243
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2023
Supplement Date Received by Manufacturer01/15/2024
Initial Report FDA Received Date11/06/2023
Supplement Report FDA Received Date11/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/04/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-