Model Number N/A
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the device had a damaged comb.There was no noted harm or delay in the procedure.During product evaluation, the cutter failed the test cut.Due diligence is complete and there is no additional information available.There was no adverse event associated with the malfunction.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record identified the following related repair: the cutter failed the test cut.The cutter will need replacement.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to design issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001526350-2023-00756-1, 0001526350-2023-01429, 0001526350-2023-01431, 0001526350-2023-01432.E1: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 4:1 RATIO ( CAUTION: SHARP )
Common Device Name
SKIN GRAFT MESHER
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18077347
Report Number0001526350-2023-01430
Device Sequence Number7729164
Product Code FZW
UDI-Device Identifier00889024375475
UDI-Public(01)00889024375475(11)180612(10)64063506
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00770304000
Device Lot Number64063506
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2023
Initial Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
00770102200ZIMMER SG MESHER RATCHET HANDL
Patient SexPrefer Not To Disclose
-
-