Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that there was a chip missing out of the broach, the device had fractured.The hospital suspects it happened during reprocessing which has caused the device to be unusable.There was no patient impact or involvement.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Tprlc 133 mp rasp 16x117mm, part # 51-206160 from lot zb161002, was returned and evaluated against the complaint.The size and lot etching on the returned device was verified to match the complaint.Visual inspection found a chip in the coating between the teeth.A review of the device history records identified no deviations or anomalies during manufacturing.Reported event did not occur in the operating room or as part of a medical procedure.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP RASP 16X117MM
Common Device Name
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18078232
Report Number0001825034-2023-02592
Device Sequence Number7823839
Product Code KWY
UDI-Device Identifier00880304516823
UDI-Public(01)00880304516823(11)161104(10)ZB161002
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K120030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-206160
Device Lot NumberZB161002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2023
Supplement Date Received by Manufacturer02/19/2024
Initial Report FDA Received Date11/06/2023
Supplement Report FDA Received Date02/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-