Catalog Number 23510070
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that during cssd washing the ¿plastic¿ section of the sleeve was found to be cracked and split.
 
Event or Problem Description
It was reported, that during cssd washing.The ¿plastic¿ section of the sleeve was found to be cracked and split.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned with a broken plastic handle/adapter.Visual inspection: the tissue protection sleeve, long imn instruments ø9mm was returned with a cracked plastic handle.This plastic handle, also presents the counterpart for the locking drill sleeve [locking mechanism of the whole sleeve assembly].Signs of damage at that area, like imprints from hitting or other kinds of mishandling were not apparent.The remaining plastic appearance was in good condition.No other deficiencies determined.Finally, although damaged, it could be assembled with returned counterpart.Functional inspection: refer to « visual inspection ».Dimensional inspection: manufacturing and inspection records confirmed, the item had been released in accordance to specs.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found, during the investigation.With given available information and since nothing was reported, during former pre-functional check or after last surgery.It could not be excluded, that¿s rather related to reprocessing steps performed.However, an exact root cause could not be determined, since individual cleaning steps remained unknown.In case substantive information will become available in future that suggests otherwise, we reserve the right to reopen the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TISSUE PROTECTION SLEEVE, LONG IMN INSTRUMENTS Ø9MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18082276
Report Number0009610622-2023-00392
Device Sequence Number20765827
Product Code LXH
UDI-Device Identifier07613327368154
UDI-Public07613327368154
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number23510070
Device Lot NumberKP477845
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2023
Supplement Date Received by Manufacturer12/20/2023
Initial Report FDA Received Date11/07/2023
Supplement Report FDA Received Date01/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-