• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON NEEDLE 25X5/8 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON NEEDLE 25X5/8 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Catalog Number 305122
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
Mat#: 305122.Batch#: 3024981.It was reported by the customer that they allege that multiple needles out of their 9cs order are leaking the needle meets the plastic portion.Customer has stopped using these units pending feedback on what to do with these units.
 
Additional Manufacturer Narrative
(b)(4): initial mdr submission.A follow up mdr will be submitted if additional information becomes available.Patient problem code: f26 ¿ no health consequences or impact.Device problem code: a0504 - leak / splash.The actual date of event is unknown.The date received by manufacturer was entered into the date of event field.
 
Additional Manufacturer Narrative
(b)(4).Follow up for device evaluation.It was reported multiple needles are leaking.To aid in the investigation, two samples with no packaging blisters were received for evaluation by our quality team.A visual inspection was performed with 10x magnification.There is a molding short shot a 5/16" from the bottom of the needle hub.Each sample was then connected to a syringe with saline solution and there was leakage from the short shot.No other defects or imperfections were observed.A device history record review was completed for provided material number 305122, lot 3024981.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.The samples will be shown to associates for awareness.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer is confirmed.
 
Event or Problem Description
(b)(4).Additional information received.  (b)(6) 2023.As indicated within the original complaint sent to you ¿ this complaint is not related to any patient or adverse effect.- please provide occurrence date.Not provided by customer.- what medication was used during the incident? not applicable.- is there any adverse event on patient? not applicable.- is there any medical intervention needed due to the incident? not applicable - is photo/sample or available for return? if yes, please confirm your address for return label.Not provided by customer.(b)(6) 2023.Please provide occurrence date.(b)(6) 2023 different days because it happens often.- what medication was used during the incident? no medication.- is there any adverse event on patient? no.- is there any medical intervention needed due to the incident? no.- is photo/sample or available for return? if yes, please confirm your address for return label.Yes i send you the photo but we also can send you some exemples.Note that we use this product to process radioactivity so it is a very big problem for us.The address to return label is 2655, av dalton , quebec g1p3s8.(b)(6) 2023 no worry, i have many samples on my desk.I will send as soon that i receive the labels.Mat#:305122 batch#:3024981.It was reported by the customer that they allege that multiple needles out of their 9cs order are leaking the needle meets the plastic portion.Customer has stopped using these units pending feedback on what to do with these units.Verbatim: rcc received the complaint via email.Email(s) as attached.Xxx alleges that multiple needles out of their 9cs order are leaking the needle meets the plastic portion.Customer has stopped using these units pending feedback on what to do with these units.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE 25X5/8 RB
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
helen cox
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18082991
Report Number1911916-2023-00826
Device Sequence Number12410061
Product Code FMI
UDI-Device Identifier30382903051220
UDI-Public(01)30382903051220
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K021475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number305122
Device Lot Number3024981
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2023
Supplement Date Received by Manufacturer11/29/2023
Initial Report FDA Received Date11/07/2023
Supplement Report FDA Received Date11/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1