• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM LUBRI-SIL® IC COMPLETE CARE®

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM LUBRI-SIL® IC COMPLETE CARE® Back to Search Results
Catalog Number A300416A
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that foley tray was missing the 10ml saline syringe.No product was saved as staff pulled missing syringe separately and used this tray.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that foley tray was missing the 10ml saline syringe.No product was saved as staff pulled missing syringe separately and used this tray.Per sample received on (b)(6) 2023, it was noted that the balloon dilation catheter was returned for evaluation.Per sample received on (b)(6) 2023, it was noted that the eagle inflation device was returned for evaluation.
 
Additional Manufacturer Narrative
The reported event is inconclusive as the reported sample was not returned for this evaluation.During visual evaluation, a box was received from the customer, but the product return (balloon dilation catheter & eagle inflation device) does not match the product in this record.Root cause could not be identified.Although a root cause could not be definitively identified, a potential root cause for this type of failure could be operator error.However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions that would have contributed to the reported event.Labelling review is not required as labelling would not have prevented the reported event.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP¿ FOLEY TRAY SYSTEM LUBRI-SIL® IC COMPLETE CARE®
Common Device Name
FOLEY TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18084539
Report Number1018233-2023-07924
Device Sequence Number12695069
Product Code EZL
UDI-Device Identifier00801741073762
UDI-Public(01)00801741073762
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue NumberA300416A
Device Lot NumberNGHT2381
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2023
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date11/07/2023
Supplement Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-