| Catalog Number |
06407358190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number of the customer's cobas 8000 c702 module is 15c7-08.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable mg2 magnesium gen.2 results from five patient samples tested on the cobas 8000 c702 module.It is unknown if the initial results were reported outside of the laboratory.The reporter suspects imprecision.The reporter was able to provide the following examples of discrepant results: (b)(6).The patient sample was run on line 2 of the module.The initial result was 0.68 mmol/l.The first repeat result was 0.47 mmol/l.The second repeat result was 0.42 mmol/l.(b)(6).The patient sample was run on line 2 of the module.The initial result was 1.15 mmol/l.The first repeat result was 0.79 mmol/l.The second repeat result was 0.81 mmol/l.(b)(6).The initial result from line 2 was 0.28 mmol/l.The first repeat result from line 2 was 0.55 mmol/l.The second repeat result from line 1 was 0.31 mmol/l.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided, the cause of the event could not be determined.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|