Model Number 0600-0783
Medical Device Problem Codes Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A replacement glidescope core smart cable was provided to the customer and the reported glidescope core smart cable used during the procedure was returned to verathon for evaluation.Upon visual inspection of the returned smart cable, the verathon technical service representative (tsr) identified hdmi connector damage and bent pins.Due to the damaged connector pins, the tsr was unable to connect the smart cable to verathon's test equipment without risk of damaging the test equipment; therefore the reported image issue could not be reproduced.Despite being unable to perform functionality testing to confirm the reported image issue, it is likely that the damaged connector pins caused or contributed to the reported issue.The glidescope video laryngoscopes operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged." verathon followed up with the customer and restated the importance of checking the device before its use in a procedure.Upon completion of verathon's device evaluation, the smart cable was scrapped due to the customer already being provided a replacement and there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event or Problem Description
A customer reported that during a patient procedure, using a glidescope core smart cable, the cable was very loose and the image would disappear intermittently when the cable was manipulated.The procedure was completed using a different glidescope core smart cable which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
 
Additional Manufacturer Narrative
This is a supplemental report correcting the unique device identifier (udi) information in section d4.Following the initial report submission, it was identified that the udi information initially provided was incorrect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18085471
Report Number9615393-2023-00199
Device Sequence Number6629155
Product Code CCW
UDI-Device Identifier00879123006752
UDI-Public(01)00879123006752(11)211213
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2023
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/07/2023
Supplement Report FDA Received Date07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-